Important: Use the tray designed for the application you select.
The imaging system detects the type of sample tray on the imaging stage. If the tray in the transilluminator drawer does not support the selected application, an error message appears and the image is not acquired.
The imaging system supports the following trays:
|
Sample Tray Name |
Icon |
|---|---|
|
Blot/UV/Stain-Free tray |
|
|
White tray |
|
|
Blue tray |
|
Some fluorescent reagents have excitation peaks in both UV and blue wavelengths. In this case, either excitation source can be used. However, the resulting images are not identical. Using the recommended tray with these fluorescent reagents provides a better signal-to-noise ratio, which increases sensitivity.
When two tray types can be used, both are shown. An asterisk identifies the recommended tray.
| 1. | Place the sample face up and center it on the appropriate sample tray. |
Tip: Center the sample carefully so it fills the image area. If you enlarge the image by zooming in, check to ensure that the sample is still centered.
| 2. | Open the imager door and pull out the transilluminator drawer. |
| 3. | Place the sample tray on the imaging stage. |
| 4. | Close the transilluminator drawer and the door. |